Features & Specifications
- Eyelet for easy mounting with cable ties
- 10-bit resolution
- User-selectable sampling intervals: 1
second to 18 hours
- High and low alarm-indication LEDs
(Pendant Temp only) — user settable levels
- Programmable Start time/date or Trigger
Start on-site with magnet
- Water-resistant housing for underwater
use — depths to 30 meters (100 feet)
- 8K or 64K memory options — store
approximately 6.5K or 52K readings
- Battery level indication at launch
- Battery level can be recorded as separate
channel
- Numerous event types (host PC connection,
low battery, magnet presence) stored to data file during
logging
- Data download via Optic USB (IR) Base
Station and Pendant Coupler
- Offload data or check logger status while
logging
- NIST-traceable temperature accuracy
certification available
- Logger size: 5.8 x 3.3 x 2.3 cm (2.3 x
1.3 x 0.9")
- Weight: 18 grams (0.6 oz)
- Patented technology (U.S. Patent
6,826,664)
HOBO Pendant Temperature Data Logger
Part # UA-001-xx
Temperature logger with high and low alarm LEDs
Key Specifications
Temperature
• Range: -20° to 70°C
(-4° to 158°F) in air; 0 to 50°C (32° to 122°F) in water
• Accuracy: 0.47°C
(0.85°F) at 25°C (77°F)
HOBO Pendant Temp/Light Intensity Data
Logger
Part # UA-002-xx
2-Channel logger
Key Specifications
Temperature
• Range: -20° to 70°C
(-4° to 158°F) in air; 0 to 50°C (32° to 122°F) in water
• Accuracy: 0.47°C
(0.85°F) at 25°C (77°F)
Light Intensity
• Range: 0—30,000
lumens/square-foot
• Designed for indoor
and outdoor measurement of relative light levels
|
HOBO Pendant Logger with Optic
Base Station and Coupler
|
 |
|
|
Optic USB Base Station/Coupler
The Optic USB Base Station and Pendant
Coupler provide fast optical infrared (IR) communication for easy
logger launch and readout of HOBO Pendant loggers. The Base
Station connects via USB to a host computer, while the Coupler
attaches the logger to the Base Station for communications.
Optic USB Base Station offloads full 64K logger in 30 seconds
Note: The Optic USB Base Station is splash resistant.
Required
Software
HOBOware™ for Windows®
software is required to operate the HOBO®
Pendant loggers. Each starter kit includes software, USB interface
cable, and software manual.
|