U12 Thermocouple Logger |
|
|
|
 |
Size/Weight 5.8 x 7.4
x 2.2 cm (2.3 x 2.9 x 0.85”)
46 gms (approximately 1.6 oz) |
The new HOBO U12 J, K, S, T Thermocouple logger
with 12-bit resolution provides high-accuracy measurement over
wide temperature ranges. One logger model accepts any J, K, S, or
T thermocouple probe with a plug-in subminiature connector;
thermocouple type is selected in software at time of deployment.
The U12 Thermocouple logger includes an internal thermistor for
cold-junction compensation and recording of ambient temperature.
The U12 Thermocouple logger requires Onset’s GreenLine™
(Windows®) or HOBOware™ (Mac) software for logger
launch, offload, and data viewing.
|
Features & Specifications
- Single-channel logger accommodates four
thermocouple types (J, K, S, T)
- Use with standard thermocouple probes with
plug-in subminiature connectors
- 12-bit resolution
- 43,000 measurements memory capacity
- Automatic cold-junction compensation
- Direct USB Interface
- 1-year battery life (typical)—user-replaceable
CR-2032 lithium battery
- Battery level indication at launch
- Battery level can be recorded as a separate
channel
- Operating Range for logger: 0 to 50°C (32° to
122°F), 0-95% relative humidity, non-condensing
Detailed Specifications
| |
Temp Measurement Ranges |
Accuracy* |
Max/Min Resolution |
Type J
(Iron-Constantan) |
0° to 750°C
(32 to 1382°F) |
±2.5°C
(±4.5°F) |
0.21° at 375°C
(0.38° at 707°F) |
| |
|
|
|
Type K
(Chromel-Alumel) |
0° to 1250°C
(32° to 2282°F) |
±4.0°C
(±7.2°F) |
0.32° at 625°C
(0.58° at 1157°F) |
| |
|
|
|
Type S
(Platinum-Rhodium) |
-50° to 1760°C
(-58° to 3200°F) |
±6.0°C
(±11°F) |
0.44° at 855°C
(0.79° at 1571°F) |
| |
|
|
|
Type T
(Copper-Constantan) |
-200° to 100°C
(-328° to 212°F) |
±1.5°C
(±2.7°F) |
0.1°C at -50°C
(0.18° at -58°F) |
| |
|
|
|
| Internal Temp |
0° to 50°C
(32° to 122°F) |
±0.35°C
(±0.63°F) |
0.03° at 25°C
(0.05° at 77°F) |
|
|
|
|
| * Logger Only |
Required Software
GreenLine
Software version 1.1 (Windows®) or HOBOware™ Software (Mac)
software is required to operate the HOBO
U12 Thermocouple logger. Each starter kit includes software, USB interface